Setaflash Series 8 Closed Cup Flash Point Tester - High Temperature - 82000-2U

An automated closed cup flash point tester called the Setaflash Series 8 has a temperature range of ambient to 300°C (ambient to 572°F). The device makes use of a hot wire ignitor.

There are two ways that a flash point can function:

  • Flash
  • No Flash and Ramp

 

By using the ramp mode, which raises the sample temperature by 2°C/min, it is possible to determine a flash point that complies with IP 534 and ASTM D7236. Every 1°C up to 100°C (and every 2°C rise above 100°C) until a flash is detected or the end of the test temperature is reached, the dipping sequence is automatically initiated. A new electrical hotwire is used by the ignitor (gas flames are an option).

Configuration Options

Key Features

  • Best published precision of any flash point method
  • Suitable for unknown samples using ramp mode
  • Automatic dipping and flash detection
  • Automatic barometric pressure correction
  • Flash/No flash mode to determine whether a product will or will not flash at a specified temperature

International Regulatory Compliance Methods

  • ASTM D3278
  • ASTM D3828
  • ASTM D7236
  • ASTM D8174
  • ASTM E502
  • BS EN ISO 3679
  • BS 2000-523
  • IP 523
  • IP 524
  • IP 534